Bomb Calorimetry · Interactive Demo

Load a sample, seal the bomb, press Ignite — watch the fuse break, the sample burn, and the temperature rise. Results include fuse-wire and HNO₃ corrections.

Apparatus

IGNITION O₂ inlet stirrer thermometer
fuse wire crucible + sample stirrer thermometer water bath

Controls

Cbomb = 900 J °C⁻¹  ·  cwater = 4.184 J g⁻¹ °C⁻¹
Total heat capacity C = Cbomb + mwater·cwater = 9268 J °C⁻¹

Live readout

Temperature22.000°C
ΔT0.000°C
Elapsed0s
Temperature vs. timeT (°C)

Corrections

Total correction removed from qtotal: 42 J
Set both to 0 to see the uncorrected result.

In a real experiment the fuse correction is found by weighing the wire before and after ignition (≈15–45 J), and the HNO₃ correction by titrating the bomb washings (≈5–25 J).

Results waiting

ΔT measured—
q total = C·ΔT—
− fuse wire—
− HNO₃ formation—
q sample (corrected)—
ΔHc per gram—
ΔHc per mole—
vs accepted—

Ignite a sample to record a measurement.

Comparison of runs 0

No completed runs yet. Ignite a sample to record your first measurement.

Schematic simulation — real bomb calorimeters also correct for heat exchange with the jacket and sulfur/ash residues. Corrections shown: fuse wire (efuse) and nitric-acid formation (eHNO₃).