What this demo does
Bomb calorimetry measures the heat of combustion of a substance. A weighed sample is
placed in a steel “bomb”, pressurised with oxygen, and ignited electrically. The bomb sits in a known
mass of water; the temperature rise of the water + bomb gives the heat released.
How to use it
- Choose a sample from the dropdown (or the unknown X to test yourself).
- Set the sample mass and the water mass with the sliders.
- Set the fuse wire and HNO₃ corrections in the Corrections panel.
- Press ⚡ Ignite. The fuse heats, breaks, and the sample burns.
- Watch the temperature curve and the live readout. After ~130 s the result appears.
- Use Compare runs to line up every measurement you have made.
The calculation
Total heat capacity of the calorimeter:
C = C_bomb + m_water × c_water
Heat measured from the temperature rise (this includes the fuse and acid):
q_total = C × ΔT
Corrected heat released by the sample alone:
q_sample = q_total − e_fuse − e_HNO₃
Enthalpy of combustion per gram and per mole (reported as magnitudes; the true sign is negative
because heat is released):
ΔH_c = q_sample / m_sample and ΔH_c,m = ΔH_c × M
Why the corrections?
-
Fuse wire — the cotton or nichrome thread burns along with the sample and adds
heat. It is found by weighing the wire before and after ignition; typical values are 15–45 J.
-
HNO₃ formation — nitrogen from the air inside the bomb oxidises to nitric acid,
which releases heat. It is found by titrating the bomb washings; typical values are 5–25 J.
Set both corrections to 0 J and run a sample to see how much the uncorrected
result differs from the accepted value. The Results note reports both errors.
Reading the results
- Green — within 0.5% of the accepted value.
- Yellow — within 2% of the accepted value.
- Red — more than 2% off.
Large water masses give a smaller ΔT, so the relative precision drops. Small samples are more
sensitive to the corrections.
About the unknown
Sample X is a white crystalline solid with a molar mass around 148 g/mol. Once you have a
measurement, compare the enthalpy with the accepted list to identify it.
Simplifications
This is a teaching simulation. It ignores heat exchange with the jacket (the “cooling correction”),
sulfur and ash residues, and the small heat of the fuse compared with the sample. Real bomb
calorimeters use a calibrated standard (usually benzoic acid) to find C_bomb in the
first place.